Overview

Overview of the FPOR measurement capabilities. Unique possibility of having a single measurement fixture for the whole 10-170 GHz range.

Repeatability

Fused silica wafer (148.15 ± 0.36 um) measured 10 times each time removing the sample and inserting again. Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio. Outstanding Dk repeatability of 0.15%!

COP foil (100 ± 1 um) measured 10 times each time removing the sample and inserting again. Dk repeatability of ±0.2% is now mainly driven by the thickness uncertainty of ±1%.

Polymers

FEP teflon foil (101.7 ± 3.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

PVC foil (147.1 ± 4 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

PET foil (147.1 ± 4.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Ultra-thin films

Mylar foil (6 ± 0.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

PEEK foil (26 ± 0.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

In-plane anisotropy

Single-crystal X-cut quartz (494.6 ± 1.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Single-crystal R-cut sapphire (115.4 ± 1.2 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

High Impact Polystyrene – HIPS foil (244.8 ± 2.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Highly-resistive semiconductors

Silicon (383.3 ± 1 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Electrically-thick samples

Sapphire window (1980 ± 1 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Sapphire window (1017.5 ± 3 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.

Sapphire window (267 ± 1.5 um). Dk uncertainty bars are due to thickness uncertainty and Df uncertainty bars are due to signal-to-noise ratio.